Failures of Heating Cooling Circulators
について heating cooling circulator can be used in many chemical pharmaceuticals and laboratories. Then, how to solve the failures of the heating cooling circulator?
Failure of circulating pump
The failure of the circulating pump is generally the heat overload of the circulating pump. There are 4 main reasons for this failure:
1. The external circulation pipeline is too long. It is suggested to shorten the external circulation pipeline, reduce the bending of the external circulation pipeline, and then click the Reset button of the electric cabinet surface heat relay;
2. The circulating pipe is blocked. It is suggested to check the degree of cleanliness of the circulating pipe, install a filter, and then click the Reset button of the electric cabinet heat relay;
3. The circulation pump voltage is too low. It is suggested to measure whether the circulation pump voltage is in line with the prescribed range, improve the voltage environment, and then click the Reset button of the electric cabinet heat relay;
4. The regulating current of the heat relay is too small. It is suggested to measure the pump current, compare the heat relay range, turn up the heat relay current properly with the cross screwdriver, and then click the Reset button of the electric cabinet heat relay.
Failure of sensor
The great possibility of sensor failure is the wiring error. At this time, check the wiring in time, and we should pay attention to two kinds of situations, one is the sensor is short-circuited or the temperature is < -160 ℃, the other is the sensor is open circuited or the temperature is > 349℃.
If the sensor is short-circuited or the temperature is <-160 ℃, it may be the fault of the sensor itself. At this time, change the sensor (there is a standby sensor in the equipment line slot), if the sensor connection is wrong, it is suggested to check the connection condition (the control circuit diagram) to correct.
If the sensor is open circuited or the temperature is > 349 ℃, it should be the fault of the sensor itself. At this time, replace the sensor in time (there is a standby sensor in the equipment line slot). It is necessary to pay attention to whether the temperature is too high and beyond the range. If the equipment is out of electricity, check the power supply of the equipment.
The above two failures of the heating cooling circulator are relatively simple. I believe that you can quickly solve these failures with these tips.
関連推奨品
-
空気によって冷却されるスリラー装置の詳しい紹介
1122LNEYA冷凍は10年以上の冷却サポートアプリケーションの経験を蓄積してきました。さまざまな応用分野とサービス条件に応じて、空冷式冷凍機を含むさまざまなタイプの冷却冷凍機を開発しました。
詳細を見る -
深冷凍ベアリングなど金属ワーク用冷凍機の導入
835LNEYAの連続凍結ワーク低温ボックスは、冷却速度が速く、低温は-150℃に達することができます。ベアリングや機械部品の冷間組立に使用できます。極低温処理は、機械的特性を向上させることができます...
詳細を見る -
高温環境チャンバーの設計とテスト技術 販売
12491. Temperature selection (1) The climatic zones of the region; Whether the product is exposed to the solar radiation environment near the heating device; Surrounding air, solar radiation and ways of heat transfer from heating devices to products;...
詳細を見る -
産業用冷凍機選定の入口
1093Industrial chillers are refrigeration devices that assemble refrigerators, condensers, evaporators, expansion valves, control systems, and switch boxes on a common base or frame. According to the heat rejection method of the unit, it can be divide...
詳細を見る