Failures of Heating Cooling Circulators
について heating cooling circulator can be used in many chemical pharmaceuticals and laboratories. Then, how to solve the failures of the heating cooling circulator?
Failure of circulating pump
The failure of the circulating pump is generally the heat overload of the circulating pump. There are 4 main reasons for this failure:
1. The external circulation pipeline is too long. It is suggested to shorten the external circulation pipeline, reduce the bending of the external circulation pipeline, and then click the Reset button of the electric cabinet surface heat relay;
2. The circulating pipe is blocked. It is suggested to check the degree of cleanliness of the circulating pipe, install a filter, and then click the Reset button of the electric cabinet heat relay;
3. The circulation pump voltage is too low. It is suggested to measure whether the circulation pump voltage is in line with the prescribed range, improve the voltage environment, and then click the Reset button of the electric cabinet heat relay;
4. The regulating current of the heat relay is too small. It is suggested to measure the pump current, compare the heat relay range, turn up the heat relay current properly with the cross screwdriver, and then click the Reset button of the electric cabinet heat relay.
Failure of sensor
The great possibility of sensor failure is the wiring error. At this time, check the wiring in time, and we should pay attention to two kinds of situations, one is the sensor is short-circuited or the temperature is < -160 ℃, the other is the sensor is open circuited or the temperature is > 349℃.
If the sensor is short-circuited or the temperature is <-160 ℃, it may be the fault of the sensor itself. At this time, change the sensor (there is a standby sensor in the equipment line slot), if the sensor connection is wrong, it is suggested to check the connection condition (the control circuit diagram) to correct.
If the sensor is open circuited or the temperature is > 349 ℃, it should be the fault of the sensor itself. At this time, replace the sensor in time (there is a standby sensor in the equipment line slot). It is necessary to pay attention to whether the temperature is too high and beyond the range. If the equipment is out of electricity, check the power supply of the equipment.
The above two failures of the heating cooling circulator are relatively simple. I believe that you can quickly solve these failures with these tips.
関連推奨品
-
LNEYA、3度目の国家ハイテク企業認証を獲得
2874In November 2021, the Science and Technology Department of Jiangsu Province released the "2021 High-tech Enterprise Record Publicity List in Jiangsu Province", and LNEYA successfully passed the review. This is the third time that the company has...
詳細を見る -
LNEYA 低温サーモスタット・バス取扱説明書
1305低温サーモスタットは、石油、化学工業、電子機器、物理工学、化学工学、生物工学、医学と健康、生命科学、食品工業、物理試験、化学分析、研究などに広く使用されています。
詳細を見る -
水冷式冷凍機の-15℃における冷凍効果
864The main configurations of water-cooledchillers are: compressors, electrical components, water pumps, evaporators,condensers, high/low pressure controllers, high/low pressure gauges, LCDthermostats, expansion valves, antifreeze switches, hand v...
詳細を見る -