塩水噴霧環境チャンバーの一般的な不具合
When you encounter some common failures of salt spray environmental chambers, how to solve them? Wuxi Guanya Refrigeration Technology Co., Ltd. has brought the relevant fault tips as follows.
Salt spray environmental chamber cannot spray
Failure causes: 1. air compressor does not work; 2. main switch of the air compressor outlet does not open; 3. solenoid valve malfunctions; 4. pressure meter malfunctions or the pressure is too low; 5. electromagnetic contactor malfunctions; 6. nozzle blocks;
Handling methods: 1. open the air compressor; 2. open the main switch of the air compressor; 3.4.5. ask the after-sales department for door-to-door repair service; 6. remove the nozzle and clean it (please carefully disassemble).
The temperature of the saturation tank of the salt spray environmental chamber cannot rise to the set temperature
Failure causes: 1. the temperature of the temperature controller of the saturation tank is too low; 2. the temperature of the saturation bucket safety protection switch is set too low; 3. heating system malfunctions; 4. electromagnetic relay malfunctions; 5. controller malfunctions.
Handling methods: 1. Set the temperature of the temperature controller to the required temperature; 2. Set the temperature of the safety switch to the required temperature; 3. Ask the after-sales department for door-to-door repair service.
Lack of spray amount in the salt spray environmental chamber
Failure causes: 1. spray regulator is placed too low; 2. glass filter pf the preheating tank is blocked; 3. pressure is set too low.
Handling methods: 1. raise the spray regulator; 2. clean up the glass filter; 3. adjust the pressure of the pressure valve to 1Kg/cm², the air compressor is marked with a pressure valve, adjust to the pressure to 2Kg/cm².
These are some handling methods of the salt spray environmental chamber failures brought by Wuxi LNEYA. Wuxi LNEYA has 環境チャンバー販売, the quality and the after-sales are guaranteed.
関連推奨品
-
2018 CIPM China|LNEYAから展示会への招待状が届きました。
1228The 56th (Fall 2018) National Pharmaceutical Machinery and China International Pharmaceutical Machinery Expo will be held from November 5th to November 7th, 2018 at Wuhan International Expo Center. As one of the professional communication platfor...
詳細を見る -
Attention points for the use of laboratory chiller manufacturers’ equipment
900実験室では、さまざまな実験器具が広く使われている。実験器具を稼働させる場合、高温低温循環装置冷凍加熱装置とは切っても切れない関係にある。何をすべきか...
詳細を見る -
暖房・冷房用サーキュレーターの選択に関する要因は何ですか?
981加熱・冷凍サーキュレーターは、多くの実験室プロジェクトで使用されている。それらは通常、冷却、加熱、温度制御のための反応器やリアクターを備えている。では、どのような要因が加熱・冷却装置の選択に関係しているのでしょうか。
詳細を見る -
医薬品製造リアクター用統合温度制御ユニット
891リアクターシステムは、統合温度制御システムと効果的に協力し、さまざまな化学反応や生産プロセスの温度制御要件を満たすことができます。反応器温度制御装置...
詳細を見る