First wipe the machine shell from top to bottom with soapy water or disinfectant commonly used in the laboratory, then wipe it thoroughly with clean water, and dry it with a soft dry cloth.
Cautions: Avoid using too much water to clean the control board area, in order to prevent leakage damage of circuit components.
Cleanenvironment of ultra-lowtemperature freezer: Once a month
Clean up all ground debris and dust with sanitary ware, reduce dust adsorption in compressor and refrigeration system, affect heat dissipation
Prevent dust from adsorbing on circuit board, causing circuit components to be damaged by interference
Sealing test of door seals:
Once every six months, poor sealing will increase the working time of compressor and cause compressor failure.
Open the door, then put in a sheet of A4 paper, close the door, pull A4 paper out, if it is difficult to pull, it shows good sealing. It needs to be tested all around the door seal.
Clean air filter: Clean at least 4 times a year
Grasp the button at the bottom left corner and open the door directly below the cryogenic freezer.
Remove the black air filter and wash it with water or neutral detergent.
After drying, place the filter in place.
If it’s already dirty, the dust and debris on it can’t be removed, and a new filter needs to be replaced.
Clean condenser: at least twice a year
Refrigerator radiates heat to the outside environment through condenser.
Ash accumulation in condenser and filter will cause heat dissipation obstruction and decrease refrigeration efficiency of refrigerator.
1. Grasp the button at the bottom left corner and open the door directly below the refrigerator.
2. Use a vacuum cleaner to clean the dust on the condenser, or use compressed air to blow off the dust adsorbed on the condenser, or brush off the dust with a fine brush.
Notes: According to the environmental conditions of different laboratories, it is necessary to clean the condenser several times a year. And in the cleaning process, do not damage the radiator on the condenser.
Intracavity defrosting: once a year or when the thickness of defrosting exceeds 10 mm
High humidity, frequent door opening can easily lead to frost inside the door. Excessive frost, ice will affect the performance of refrigerators, can be deicing shovel and vacuum cleaner deicing.
1. Transfer samples from all chambers to another cryogenic refrigerator
2. Turn off the battery switch, the main power switch, and pull out the power cord.
3. Open all the inner doors and let the ice melt and wipe it off with a soft rag.
4. When the frost is removed, the inner cavity needs to be thoroughly cleaned with hot water or detergent (soda water) without chloride and wiped with a dry cloth.
5. Re-insert the power cord and turn on the main power switch and battery switch.
6. Samples can be placed after the refrigerator runs empty overnight.
我々は 超低温冷凍庫メーカー.私達のプロダクトの何れかのモデルに興味がある場合は、お問い合わせは、自由にしてください a>を参照してください。
関連推奨品
-
LNEYA 高温・低温加熱式冷蔵庫 メンテナンス手順書
1158循環水を使用する高温・低温加熱冷蔵庫は、循環水を1カ月ごとに定期的に交換する。熱媒油を使用している場合は、半年から1年ごとに熱媒油を定期的に交換する。高温・低温加熱式冷蔵庫の場合
詳細を見る -
なぜ半導体製造企業は加熱循環装置を必要とするのか?
1077温度制御の精度:半導体製造プロセスでは、温度制御に対する要求が非常に厳しくなっています。加熱サーキュレーターは正確な温度制御が可能であり、半導体製造工程で使用される温度は...
詳細を見る -
ベアリング低温処理装置の利点
1121超低温フリーザーの名前の一つはベアリング極低温処理装置です。無錫莞亜LNEYAの超低温フリーザーは、大容積比、大容量ベアリング容量設計、温度を設定することができます。
詳細を見る -
Low temperature chilled lubricant selection considerations
1257When you use a cryogenic chilled, you definitely want to know about all aspects of the cryocooler. But one of the things that needs everyone's attention is that the type of lubricant for the chilled is chosen because there are many influencing fac...
詳細を見る